Development of an Infrared Fluorescent Gas Analyzer.
McClatchie, E. A.
A prototype model low level carbon monoxide analyzer was developed using fluorescent cell and negative chopping techniques to achieve a device superior to state of art NDIR (Nondispersive infrared) analyzers in stability and cross-sensitivity to other gaseous species. It is clear that this type of analyzer has that capacity. The prototype analyzer, which is a self-contained unit powered from 115V 60HZ line, showed an ultimate detectivity to CO of approximately 1 ppm. and no measurable cross sensitivity to any other gaseous species at the level normally found in the ambient atmosphere. The theory and results of the research are summarized. (NTIS)
National Technical Information Service, Springfield, Virginia 22151 (PB-213-846/9, PC-$3.00, MF-$2.25)
Publication Type: Reports - Research
Education Level: N/A
Audience: N/A
Language: N/A
Sponsor: Environmental Protection Agency, Research Triangle Park, NC.
Authoring Institution: Akron Scientific Labs., Berkeley, CA.