Reliability and Minimal Detectable Change of the Styku 3D Body Scanner
Silver, Blair; Wilson, Patrick B.
Measurement in Physical Education and Exercise Science, v24 n3 p165-171 2020
The objective of this study was to assess the reliability and minimal detectable change (MDC) of the Styku S100 3-dimensional (3D) scanner. Thirty-three participants completed two visits separated by 1-3 days; at each visit, two scans were performed. Intra-class correlation co-efficients (ICCs) and MDC values based on a 90% confidence interval (MDC90) were calculated. For body fat percentage, within-day ICC and MDC[subscript 90] values were 0.997 and 1.1%, respectively, and these values changed to 0.989 and 2% when using single scans on separate days. For 21 circumference sites, 10 of the within-day ICCs were [greater than or equal to] 0.99 and 20 of the MDC[subscript 90] values were < 2 cm. Comparatively, five ICCs were [greater than or equal to] 0.99 and 15 MDC[subscript 90] values were < 2 cm when using single scans on separate days. In conclusion, between-visit differences of 1-2% for body fat percentage and 1-2 cm for most circumference values are needed to exceed typical measurement error.
Routledge. Available from: Taylor & Francis, Ltd. 530 Walnut Street Suite 850, Philadelphia, PA 19106. Tel: 800-354-1420; Tel: 215-625-8900; Fax: 215-207-0050; Web site: http://www.tandf.co.uk/journals
Publication Type: Journal Articles; Reports - Research
Education Level: N/A
Audience: N/A
Language: English
Sponsor: N/A
Authoring Institution: N/A
Grant or Contract Numbers: N/A
Author Affiliations: N/A